- W2014799394 abstract "In Brief Study Design. Finite element model of L3–S1 segment and confirmatory cadaveric testing were used to investigate the biomechanical effects of a mobile core type artificial disc (Charité artificial disc; DePuy Spine, Raynham, MA) on the lumbar spine. Objective. To determine the effects of the Charité artificial disc across the implanted and adjacent segments. Summary of Background Data. Biomechanical studies of artificial discs that quantify parameters, like the load sharing and stresses, are sparse in the literature, especially for mobile-type core artificial disc designs. In addition, there is no standard protocol for studying the adjacent segmental effects of such implants. Methods. Human osteo-ligamentous spines (L1–S1) were tested before and after L5–S1 Charité artificial disc placement. The data were used to validate further an intact 3-dimensional (3-D) nonlinear L3–S1 finite element model. The model was subjected to 400-N axial compression and 10.6 Nm of flexion/extension pure moments (load control) or pure moments that produced the overall rotation of the L3–S1 Charité model equal to the intact case (hybrid approach). Resultant motion, load, and stress parameters were analyzed at the experimental and adjacent levels. Results. Finite element model validation was achieved only with the load-controlled experiments. The hybrid approach, believed to be more clinically relevant, revealed that Charité artificial disc leads to motion increases in flexion (19%) and extension (44%) at the L5–S1 level. At the instrumented level, the decrease in the facet loads was less than at the adjacent levels; the corresponding decrease being 26% at L3–L4, 25% at L4–L5, and 13.4% at L5–S1 when compared to the intact. Intradiscal pressure changes in the L4–L5 and L3–L4 segments were minimal. Shear stresses at the Charité artificial disc-L5 endplate interface were higher than those at S1 interface. However, in the load control mode, the increase in facet loads in extension was approximately 14%, as compared to the intact case. Conclusions. The hybrid testing protocol is advocated because it better reproduces clinical observations in terms of motion following surgery, using pure moments. Using this approach, we found that the Charité artificial disc placement slightly increases motion at the implanted level, with a resultant increase in facet loading when compared to the adjacent segments, while the motions and loads decrease at the adjacent levels. However, in the load control mode that we believe is not that clinically relevant, there was a large increase in motion and a corresponding increase in facet loads, as compared to the intact. An experimentally validated finite element model and an in vitro experimental study were pursued to assess the effects of the Charité artificial disc (DePuy Spine, Raynham, MA) on segmental biomechanics. A novel hybrid loading protocol, thought to be more clinically relevant, showed less motion perturbation at the implanted level than a classic flexibility protocol. The highest increase was in extension motion and the corresponding facet loading at the implanted level and a decrease across adjacent segments." @default.
- W2014799394 created "2016-06-24" @default.
- W2014799394 creator A5004559115 @default.
- W2014799394 creator A5007701933 @default.
- W2014799394 creator A5026610320 @default.
- W2014799394 creator A5040410550 @default.
- W2014799394 creator A5054625823 @default.
- W2014799394 creator A5057891490 @default.
- W2014799394 creator A5062454278 @default.
- W2014799394 creator A5066130648 @default.
- W2014799394 creator A5072967959 @default.
- W2014799394 creator A5076720701 @default.
- W2014799394 creator A5080886228 @default.
- W2014799394 creator A5081629487 @default.
- W2014799394 creator A5087052628 @default.
- W2014799394 date "2005-12-01" @default.
- W2014799394 modified "2023-10-10" @default.
- W2014799394 title "Effects of Charité Artificial Disc on the Implanted and Adjacent Spinal Segments Mechanics Using a Hybrid Testing Protocol" @default.
- W2014799394 cites W1964994349 @default.
- W2014799394 cites W1971438663 @default.
- W2014799394 cites W1981286572 @default.
- W2014799394 cites W1981322555 @default.
- W2014799394 cites W1981927924 @default.
- W2014799394 cites W2007712962 @default.
- W2014799394 cites W2009139761 @default.
- W2014799394 cites W2010117675 @default.
- W2014799394 cites W2012269225 @default.
- W2014799394 cites W2017564666 @default.
- W2014799394 cites W2022774761 @default.
- W2014799394 cites W2023155338 @default.
- W2014799394 cites W2024951836 @default.
- W2014799394 cites W2025034601 @default.
- W2014799394 cites W2032001292 @default.
- W2014799394 cites W2048437736 @default.
- W2014799394 cites W2051878668 @default.
- W2014799394 cites W2058662509 @default.
- W2014799394 cites W2063552028 @default.
- W2014799394 cites W2069054539 @default.
- W2014799394 cites W2071763312 @default.
- W2014799394 cites W2079204812 @default.
- W2014799394 cites W2081207873 @default.
- W2014799394 cites W2085871791 @default.
- W2014799394 cites W2093473264 @default.
- W2014799394 cites W2100317458 @default.
- W2014799394 cites W2121684649 @default.
- W2014799394 cites W2134916765 @default.
- W2014799394 cites W2405057598 @default.
- W2014799394 cites W2410527128 @default.
- W2014799394 doi "https://doi.org/10.1097/01.brs.0000195897.17277.67" @default.
- W2014799394 hasPubMedId "https://pubmed.ncbi.nlm.nih.gov/16371899" @default.
- W2014799394 hasPublicationYear "2005" @default.
- W2014799394 type Work @default.
- W2014799394 sameAs 2014799394 @default.
- W2014799394 citedByCount "203" @default.
- W2014799394 countsByYear W20147993942012 @default.
- W2014799394 countsByYear W20147993942013 @default.
- W2014799394 countsByYear W20147993942014 @default.
- W2014799394 countsByYear W20147993942015 @default.
- W2014799394 countsByYear W20147993942016 @default.
- W2014799394 countsByYear W20147993942017 @default.
- W2014799394 countsByYear W20147993942018 @default.
- W2014799394 countsByYear W20147993942019 @default.
- W2014799394 countsByYear W20147993942020 @default.
- W2014799394 countsByYear W20147993942021 @default.
- W2014799394 countsByYear W20147993942022 @default.
- W2014799394 countsByYear W20147993942023 @default.
- W2014799394 crossrefType "journal-article" @default.
- W2014799394 hasAuthorship W2014799394A5004559115 @default.
- W2014799394 hasAuthorship W2014799394A5007701933 @default.
- W2014799394 hasAuthorship W2014799394A5026610320 @default.
- W2014799394 hasAuthorship W2014799394A5040410550 @default.
- W2014799394 hasAuthorship W2014799394A5054625823 @default.
- W2014799394 hasAuthorship W2014799394A5057891490 @default.
- W2014799394 hasAuthorship W2014799394A5062454278 @default.
- W2014799394 hasAuthorship W2014799394A5066130648 @default.
- W2014799394 hasAuthorship W2014799394A5072967959 @default.
- W2014799394 hasAuthorship W2014799394A5076720701 @default.
- W2014799394 hasAuthorship W2014799394A5080886228 @default.
- W2014799394 hasAuthorship W2014799394A5081629487 @default.
- W2014799394 hasAuthorship W2014799394A5087052628 @default.
- W2014799394 hasConcept C105702510 @default.
- W2014799394 hasConcept C127413603 @default.
- W2014799394 hasConcept C135628077 @default.
- W2014799394 hasConcept C136229726 @default.
- W2014799394 hasConcept C15744967 @default.
- W2014799394 hasConcept C159985019 @default.
- W2014799394 hasConcept C170700871 @default.
- W2014799394 hasConcept C175696284 @default.
- W2014799394 hasConcept C180016635 @default.
- W2014799394 hasConcept C187288502 @default.
- W2014799394 hasConcept C192562407 @default.
- W2014799394 hasConcept C2775871850 @default.
- W2014799394 hasConcept C2865642 @default.
- W2014799394 hasConcept C43122875 @default.
- W2014799394 hasConcept C44575665 @default.
- W2014799394 hasConcept C66938386 @default.
- W2014799394 hasConcept C71924100 @default.
- W2014799394 hasConcept C77805123 @default.