Matches in SemOpenAlex for { <https://semopenalex.org/work/W2112785780> ?p ?o ?g. }
Showing items 1 to 61 of
61
with 100 items per page.
- W2112785780 endingPage "S789" @default.
- W2112785780 startingPage "S789" @default.
- W2112785780 abstract "Purpose/Objective(s)Intensity-Modulated Radiosurgery (IMRS) was adopted in our clinic for treating intracranial lesions since 2006. RapidArc (RA), a modulated technique of rotational beams, was implemented in 2009. This study reports the comparison in dosimetric and delivery aspects of the two techniques in treating multiple brain metastases.Materials/MethodsTwelve patients with 2 to12 brain metastases treated by IMRS were re-planned with RA. The IMRS plans were delivered using a single-isocenter, and in a single fraction using 9-11 static beams and 6-9 couch angles. The prescribed dose was 18-20 Gy covering 98% of the PTV. Two coplanar arcs were adopted in all RA plans. The RA plans followed systematic strategies in arc optimization. Two or more RA plans were generated for each patient and the one with lower MUs and higher conformity was chosen for comparison with IMRS. For dosimetric evaluation, the homogeneity, conformity, and gradient indices were calculated and compared for each patient. The homogeneity index is defined as the ratio of the maximum PTV dose and the prescribed dose. The conformity index is the ratio of 100% dose volume and the PTV. The gradient index is the radii difference of two equivalent spheres of the volumes enclosed by 100% and 50% doses.ResultsIt is found that MU reduction by RA is more pronounced for patients with a larger number of brain lesions, as the ratio of MUs in RA and IMRS is reduced from 104% to 39% as the lesion increases from 2 to 12. It is also found that 2-arcs gave a better dose conformity than 1-arc but with comparable MUs. Adding vertical or oblique arcs actually increased the dose to the normal brain and brainstem with little benefit in dose conformity. The homogeneity indices in IMRS and RA are found to be comparable in each patient and the group average is 1.15 for both techniques. The conformity and gradient indices are slightly higher in RA, with an average conformity index of 1.6 and 1.3 and a gradient index of 2.3 cm and 2.0 cm, respectively. The mean dose to the normal brain, and maximal dose to brainstem, chiasm, eyes, and optic nerves were calculated in each plan and compared for each patient. It is found that the dosimetric data of those organs at risk are comparable but RA plans give slightly higher doses.ConclusionsRA has pros and cons in the radiosurgery of brain metastasis. Treatment time reduction is more significant in brain radiosurgery, since RA uses less MUs, fewer beams and couch angles than IMRS technique. Homogeneities are comparable in RA and IMRS, but RA results in slightly higher conformity and gradient indices, and critical organ doses. This may be due to increased leakage and transmission through the multileaf collimator, as larger field sizes are used for multiple brain lesions in RA, while beams in IMRS are designed to avoid critical organs as much as possible. Purpose/Objective(s)Intensity-Modulated Radiosurgery (IMRS) was adopted in our clinic for treating intracranial lesions since 2006. RapidArc (RA), a modulated technique of rotational beams, was implemented in 2009. This study reports the comparison in dosimetric and delivery aspects of the two techniques in treating multiple brain metastases. Intensity-Modulated Radiosurgery (IMRS) was adopted in our clinic for treating intracranial lesions since 2006. RapidArc (RA), a modulated technique of rotational beams, was implemented in 2009. This study reports the comparison in dosimetric and delivery aspects of the two techniques in treating multiple brain metastases. Materials/MethodsTwelve patients with 2 to12 brain metastases treated by IMRS were re-planned with RA. The IMRS plans were delivered using a single-isocenter, and in a single fraction using 9-11 static beams and 6-9 couch angles. The prescribed dose was 18-20 Gy covering 98% of the PTV. Two coplanar arcs were adopted in all RA plans. The RA plans followed systematic strategies in arc optimization. Two or more RA plans were generated for each patient and the one with lower MUs and higher conformity was chosen for comparison with IMRS. For dosimetric evaluation, the homogeneity, conformity, and gradient indices were calculated and compared for each patient. The homogeneity index is defined as the ratio of the maximum PTV dose and the prescribed dose. The conformity index is the ratio of 100% dose volume and the PTV. The gradient index is the radii difference of two equivalent spheres of the volumes enclosed by 100% and 50% doses. Twelve patients with 2 to12 brain metastases treated by IMRS were re-planned with RA. The IMRS plans were delivered using a single-isocenter, and in a single fraction using 9-11 static beams and 6-9 couch angles. The prescribed dose was 18-20 Gy covering 98% of the PTV. Two coplanar arcs were adopted in all RA plans. The RA plans followed systematic strategies in arc optimization. Two or more RA plans were generated for each patient and the one with lower MUs and higher conformity was chosen for comparison with IMRS. For dosimetric evaluation, the homogeneity, conformity, and gradient indices were calculated and compared for each patient. The homogeneity index is defined as the ratio of the maximum PTV dose and the prescribed dose. The conformity index is the ratio of 100% dose volume and the PTV. The gradient index is the radii difference of two equivalent spheres of the volumes enclosed by 100% and 50% doses. ResultsIt is found that MU reduction by RA is more pronounced for patients with a larger number of brain lesions, as the ratio of MUs in RA and IMRS is reduced from 104% to 39% as the lesion increases from 2 to 12. It is also found that 2-arcs gave a better dose conformity than 1-arc but with comparable MUs. Adding vertical or oblique arcs actually increased the dose to the normal brain and brainstem with little benefit in dose conformity. The homogeneity indices in IMRS and RA are found to be comparable in each patient and the group average is 1.15 for both techniques. The conformity and gradient indices are slightly higher in RA, with an average conformity index of 1.6 and 1.3 and a gradient index of 2.3 cm and 2.0 cm, respectively. The mean dose to the normal brain, and maximal dose to brainstem, chiasm, eyes, and optic nerves were calculated in each plan and compared for each patient. It is found that the dosimetric data of those organs at risk are comparable but RA plans give slightly higher doses. It is found that MU reduction by RA is more pronounced for patients with a larger number of brain lesions, as the ratio of MUs in RA and IMRS is reduced from 104% to 39% as the lesion increases from 2 to 12. It is also found that 2-arcs gave a better dose conformity than 1-arc but with comparable MUs. Adding vertical or oblique arcs actually increased the dose to the normal brain and brainstem with little benefit in dose conformity. The homogeneity indices in IMRS and RA are found to be comparable in each patient and the group average is 1.15 for both techniques. The conformity and gradient indices are slightly higher in RA, with an average conformity index of 1.6 and 1.3 and a gradient index of 2.3 cm and 2.0 cm, respectively. The mean dose to the normal brain, and maximal dose to brainstem, chiasm, eyes, and optic nerves were calculated in each plan and compared for each patient. It is found that the dosimetric data of those organs at risk are comparable but RA plans give slightly higher doses. ConclusionsRA has pros and cons in the radiosurgery of brain metastasis. Treatment time reduction is more significant in brain radiosurgery, since RA uses less MUs, fewer beams and couch angles than IMRS technique. Homogeneities are comparable in RA and IMRS, but RA results in slightly higher conformity and gradient indices, and critical organ doses. This may be due to increased leakage and transmission through the multileaf collimator, as larger field sizes are used for multiple brain lesions in RA, while beams in IMRS are designed to avoid critical organs as much as possible. RA has pros and cons in the radiosurgery of brain metastasis. Treatment time reduction is more significant in brain radiosurgery, since RA uses less MUs, fewer beams and couch angles than IMRS technique. Homogeneities are comparable in RA and IMRS, but RA results in slightly higher conformity and gradient indices, and critical organ doses. This may be due to increased leakage and transmission through the multileaf collimator, as larger field sizes are used for multiple brain lesions in RA, while beams in IMRS are designed to avoid critical organs as much as possible." @default.
- W2112785780 created "2016-06-24" @default.
- W2112785780 creator A5000950305 @default.
- W2112785780 creator A5003454331 @default.
- W2112785780 creator A5005962112 @default.
- W2112785780 creator A5032912945 @default.
- W2112785780 creator A5050647874 @default.
- W2112785780 creator A5070663976 @default.
- W2112785780 date "2010-11-01" @default.
- W2112785780 modified "2023-09-27" @default.
- W2112785780 title "Comparison Study of Rotational vs. Static Intensity Modulated Radiosurgery for Multiple Brain Metastases" @default.
- W2112785780 doi "https://doi.org/10.1016/j.ijrobp.2010.07.1828" @default.
- W2112785780 hasPublicationYear "2010" @default.
- W2112785780 type Work @default.
- W2112785780 sameAs 2112785780 @default.
- W2112785780 citedByCount "0" @default.
- W2112785780 crossrefType "journal-article" @default.
- W2112785780 hasAuthorship W2112785780A5000950305 @default.
- W2112785780 hasAuthorship W2112785780A5003454331 @default.
- W2112785780 hasAuthorship W2112785780A5005962112 @default.
- W2112785780 hasAuthorship W2112785780A5032912945 @default.
- W2112785780 hasAuthorship W2112785780A5050647874 @default.
- W2112785780 hasAuthorship W2112785780A5070663976 @default.
- W2112785780 hasBestOaLocation W21127857801 @default.
- W2112785780 hasConcept C120665830 @default.
- W2112785780 hasConcept C121332964 @default.
- W2112785780 hasConcept C126838900 @default.
- W2112785780 hasConcept C2780387249 @default.
- W2112785780 hasConcept C2989005 @default.
- W2112785780 hasConcept C509974204 @default.
- W2112785780 hasConcept C71924100 @default.
- W2112785780 hasConcept C93038891 @default.
- W2112785780 hasConceptScore W2112785780C120665830 @default.
- W2112785780 hasConceptScore W2112785780C121332964 @default.
- W2112785780 hasConceptScore W2112785780C126838900 @default.
- W2112785780 hasConceptScore W2112785780C2780387249 @default.
- W2112785780 hasConceptScore W2112785780C2989005 @default.
- W2112785780 hasConceptScore W2112785780C509974204 @default.
- W2112785780 hasConceptScore W2112785780C71924100 @default.
- W2112785780 hasConceptScore W2112785780C93038891 @default.
- W2112785780 hasIssue "3" @default.
- W2112785780 hasLocation W21127857801 @default.
- W2112785780 hasOpenAccess W2112785780 @default.
- W2112785780 hasPrimaryLocation W21127857801 @default.
- W2112785780 hasRelatedWork W141524600 @default.
- W2112785780 hasRelatedWork W1841185769 @default.
- W2112785780 hasRelatedWork W1985895619 @default.
- W2112785780 hasRelatedWork W2049214470 @default.
- W2112785780 hasRelatedWork W2072926321 @default.
- W2112785780 hasRelatedWork W2144161415 @default.
- W2112785780 hasRelatedWork W2155887765 @default.
- W2112785780 hasRelatedWork W2902148150 @default.
- W2112785780 hasRelatedWork W3093986620 @default.
- W2112785780 hasRelatedWork W3210245320 @default.
- W2112785780 hasVolume "78" @default.
- W2112785780 isParatext "false" @default.
- W2112785780 isRetracted "false" @default.
- W2112785780 magId "2112785780" @default.
- W2112785780 workType "article" @default.